Home / Injections & Infusion / Medriva 0.3ml Auto-Disable Syringe with 23G x 1” Needle

Medriva

Medriva 0.3ml Auto-Disable Syringe with 23G x 1” Needle

Medriva’s WHO-PQS AD syringes are recommended for immunization programmes because they are designed to prevent re-use by locking automatically after a single use. They are the best way of ensuring that people receive their shots safely, without a risk of contracting an infection due to contaminated needles. 

UNICEF SUPPLY DIVISION

How to buy

Reach out to our team for ordering information.

Medriva’s WHO-PQS Auto-disable syringes for fixed dose immunization are safe and easy to use. The evaluated technology activates the auto-disabling function the moment the dosage has been taken up. The syringe is rendered unusable immediately after injection of the vaccine. No manual activation is required.

Medriva’s auto-disable syringes are already used for various vaccination campaigns including on the current SARS-CoV-2 vaccines.

The product is sterilized by ethylene oxide (EO) gas.

  • Auto disable syringes are used for administering medications using standard and specialised techniques
  • The transparent syringe barrel ensures controlled administration of medication
  • Smooth-glide plunger ensures smooth, painless injection
  • Secure plunger stop prevents loss of medication
  • Clearly legible graduation ensures a safe, reliable dosage
  • The product range includes two models of the auto disable syringes:
  • Auto disable syringes with a spring lock
  • Auto disable syringes with a mechanism separating the plunger seal and the plunger
  • Latex-free plunger seal eliminates the risk of allergic reactions

Auto disable syringe

Biocompatibility

The products comply with the requirements of standard for biocompatibility of medical devices, ISO 10993-1 Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process.

Sterilization

Ethylene, Oxide Sterilization following ISO 11135 Sterilization of health-care products — Ethylene oxide — Requirements for the development, validation and routine control of a sterilization process for medical devices.

Shelf life

Shelf life: 5 years

Storage

No special storage or transportation conditions. Recommendations to store in room temperature. Store in dry and warm place and not exposed to strong light.

The Description of Product Structure and Raw Material

Product StructureRaw MaterialAppearance Color
BarrelPP 5090TClear with black printing scale
PlungerPP 5090TClear
PistonLatex Free
(Polyisoprene Rubber)
Black
Protective capPP 5090TColor according to needle size
Steel clipMedical SUS 304/
NeedleMedical SUS 304/

Technical Data Sheet

PropertiesReference
Standards
UnitsValues
Totally length of assembly productISO 7886-3mm119.2mm
Minor axis length of barrel flangesISO 7886-3mm9.9±0.2
Major axis length of barrel flangesISO 7886-3mm20.2±0.25
Total length of nominal capacity scaleISO 7886-3mm17.4+0.5mm
Distance of push-buttonISO 7886-3mmNot less than 8
Maximum volume of dead spaceISO 7886-3mL0.07
Tolerance on nominal capacityISO 7886-3%5%
Force – Initial force(Fs) maxISO 7886-1N10
Force – Mean force(F) max.ISO 7886-1N5
Force – Maximum force(Fmax)ISO 7886-1N<(2.0×measured F or (measured f+1.5N,whichever is higher
Sterilization methodISO 7886-1
ISO 11135
Sterilized by EO
EO sterilization residualsISO10993-7
GB15810
μ g/g≤10

No downloads

Medriva is built on protection. We strive to achieve patient health. And we’ll get there using innovation, product development, and the most advanced R&D team.